The Nimbus 4 system provides effective active pressure redistribution
(alternating therapy) combined with automatic adjustment to the weight, size and position of the individual. It combines Wound Valve Technology, Heelguard, and automatic adjustment of cell pressures, providing effective pressure redistribution to a wide range of patient management issues.
Nimbus range pump features
Audible and visual alarms
• The pump includes audible and visual alarms, activating only when genuine intervention is needed for Power Fail, High or Low Pressure and Pump Fault conditions.
Comfort control
• For additional patient comfort, minor adjustments to inflation pressures can be made using a comfort control dial, without affecting therapeutic performance.¹⁶
Infection prevention
• 3 stage filtration provided by the bio-filter ensures only clean air is passed into the mattress.
Nimbus range mattress features
Cover
• Vapor permeable and water resistant 2-way stretch cover offers pressure redistribution and comfort. This loose fitting top cover is attached to the mattress base by concealed zippers to reduce ingress of unwanted fluids. The cover is also flame retardant, easy to clean and can be completely removed for laundering.
Auto-matt sensor pad
• The Auto-Matt sensor pad is designed to adjust cell pressures according to patient weight and positioning on the mattress.
Anti-sink mattress design
• The mattress is made up of 19 deep "figure of 8" designed cells to support patients up to 550 lbs (250 kg), providing active pressure redistribution for all patient types and all categories⁹ of tissue damage.
Wound valve technology
• 5 heel cells incorporate Wound Valve technology, allowing the carer to permanently off-load ("float") the vulnerable heel, for those patients with limb ischemia or existing wounds.
Cable management
• Reduces risk of injury to patient and/or carer by removing trailing cables.
Transport facility
• Seals air in the mattress for up to 12 hours, providing a uniform stable surface for patient transportation, moving and handling, nursing procedures and during periods of power failure.
Heelguard
• Dedicated mattress zone where bottom 5 cells provide maximum pressure redistribution through use of unique "power down" straps, keeping pressure at the vulnerable heel lower for longer (Nimbus 4).
CPR facility
• A quick release, single handed operation, the CPR facility enables rapid air deflation for emergencies, short term physiotherapy, moving and handling or storage.
Nimbus Professional additional features
Zoned wound valve technology
• Zoned Wound Valve Technology allows carers to selectively isolate any one of 19 individual cells under a patient’s body, providing an adaptable support surface for the management of highly vulnerable areas such as grafts, burns, heels and ischemic legs.¹, ⁴, ¹³, ¹⁴
Sacral deflate
• Wound Valves in the sacral section can deflate the mattress to assist with nursing procedures including bed-chair transfer, physiotherapy, low-height patient egress and also specialized interventions such as diagnostic imaging.
Head section deflate
• The 3 head cells can be deflated to remove pressure from beneath the head, or for providing access to the head and neck for specialist nursing procedures (e.g. intubation, cannulation and hygiene).
- Audible and visual alarms
- Comfort control
- Infection prevention
- Auto-matt sensor pad
- Anti-sink mattress design
- Wound valve technology
- Cable management
- Transport facility
- Heelguard
- CPR facility
- Zoned wound valve technology
- Sacral deflate
- Head section deflate
Dimensions | |
---|---|
Mattress A Name | Standard |
Mattress A (L x W x H) mm | 2085 x 890 x 215 mm |
Mattress A (L x W x H) in | 82 x 35 x 8.5 in |
Mattress A Weight kg | 15,5 kg |
Mattress B Name | Narrow |
Mattress B (L x W x H) mm | 2085 x 800 x 215 mm |
Mattress B (L x W x H) in | 82 x 31.5 x 8.5 in |
Mattress B Weight kg | 14,3 kg |
Mattress B Weight lb | 31,5 lb |
Maximum patient weight kg | 250 kg |
Maximum patient weight lb | 550 lb |
Cell Material | Polyurethane |
Pump | |
Pump (L x W x H) mm | 508 x 220 x 100 mm |
Pump (L x W x H) in | 20 x 8.75 x 4 in |
Weight kg | 5,7 kg |
Weight lb | 12,5 lb |
Mode of Operation | Continuous |
Electrical Rating | 35VA |
Operating Cycle Time | 10 minutes |
Shock Protection | Class II Type BF |
Ingress Protection | IP21 |
Power In | 230 VAC, 50Hz |
Standards | EN60601-1, IEC60601-1 |
18 Month Non-Prorated.